How to Adjust the Comprehensive Dental Chair Treatment Machine?

The headrest cushion, backrest cushion, and seat cushion of the dental chair treatment machine should be relatively soft and comfortable, and have elasticity. When installed, special working conditions such as high-speed turbine dental drill phones should be considered. The rinsing water supply device on the dental chair is also a very important device that is used countless times during the tooth examination process. Its requirement is to have a very soft and direct water output without splashing everywhere. Otherwise, it will splash the patient and cause unnecessary quarrels, affecting their mood. The dental lamp on the dental chair treatment machine should be able to move freely and flexibly on the lamp bracket and stay stable in the required position without shaking back and forth. It will interfere with the doctor's diagnosis. When it is not subjected to external forces, the light intensity of the dental lamp cannot be changed casually.


Before using the dental chair treatment machine, it is necessary to check its levelness


Place the instrument tray in front of the machine center and use a level measuring instrument to measure the instrument tray's deviation angle. It should meet the requirements for normal use. Pay attention to the rinsing water supply device that needs to be used during the patient examination. Its outlet water injection should be checked by placing it in a glass on the rinsing cup holder. Based on visual inspection, the water condition can be observed. Only if it meets the requirements, can it be operated.


Adjustment method of the dental chair treatment machine


Check the power supply 


Press the up, down, forward, and backward keys, and the chair does not move. Remove the front cover under the patient's chair and check that the pneumatic switch of the control circuit is not closed, resulting in the control circuit not being connected. The dental chair manufacturer analyzed the reason: the screw fixing the pneumatic switch was loose, and the air valve could not effectively press the travel switch. Solution: readjust the distance between the pneumatic control valve and the travel switch to ensure that the travel switch can reliably connect under the rated air pressure, and tighten the screws.


Press the up, down, forward, and backward keys of the patient chair, and you can hear the relay closing sound, but the chair does not move


This phenomenon indicates that the control circuit is intact, check whether AC24V and AC12V are normal, in which AC12V provides power to the control circuit and AC24V provides power to the DC motor. Check the rectification voltage and find that there is no DC24V output. Looking at the circuit control panel, it was found that there was a scorch mark around the DC24V rectifier bridge. According to measurement, two arms of the rectifier bridge were damaged. Although the relay was turned off, the chair did not work because the motor did not receive electricity due to the lack of DC24V power supply. The comprehensive dental chair treatment machine manufacturer analyzed the reasons: firstly, the quality of the components was not up to standard; secondly, the motor was stuck or short-circuited. Continue to check whether the motor is normal, and exclude the second reason. Solution: Since the printed circuit around the rectifier bridge on the circuit control board has been burnt out, it is difficult to replace only the rectifier bridge, so it is decided to replace the entire circuit board. When replacing the new circuit board, the 1203 chip on the original circuit board should be removed and installed on the new board to continue using the data. The removed wires, connectors, etc. should be connected according to the original position and color code. After power-on debugging, the dental chair works normally.


Press the up, down, forward, and backward keys of the patient chair, and the chair does not move, and there is no sound of the relay closing


Open the front cover, check that AC and DC power supplies are normal, all plugs and connections are not loose, and the pneumatic switch is properly connected. At this time, check whether the signal wire of the key is normal. Check: unplug the signal wire plug on the circuit control board, connect one end of a wire to the black line socket, and touch other electrodes on the socket to observe whether the relay operates and whether the corresponding indicator light is on. If it is normal, there is a problem with the signal wire.


The treatment machine on the integrated dental chair integrates the three major elements of gas, water, and electricity in the oral treatment process. The comprehensive dental chair treatment machine manufacturer has borrowed design experience from previous products, referred to the design solutions of advanced products, and determined a design concept based on convenience, availability, and reliability during the design process.

Popular Dental Supplies

Related AJAX News

How to Adjust the Comprehensive Dental Chair Treatment Machine?
AJAX Medical Equipment Co., Ltd. is a professional dental equipment manufacturer specialilzing in the research & development, manufacturing and sales of high-end dental equipment.
Contact Us
info@ajaxdent.com
0086 020 84847938
Building No.2, Dagang Industrial Zone, Shilou Town, Panyu District, Guangzhou, P.R. China
Guangzhou Ajax Medical Equipment Co,.Ltd.
AJAX Medical Equipment Co., Ltd. is a professional dental equipment manufacturer specialilzing in the research & development, manufacturing and sales of high-end dental equipment.
Building No.2, Dagang Industrial Zone, Shilou Town, Panyu District, Guangzhou, P.R. China
Email
info@ajaxdent.com
Call Us
0086 020 84847938